Del Mar Photonics - Newsletter Fall 2010 - Newsletter Winter 2010
Del Mar Photonics supplies multi-photon lasers and systems based on cost effective femtosecond sources:
Multiphoton Imaging
Related Del Mar Photonics products
If you can't find information about the products that you are looking for just e-mail our Sales Team and we'll e-mail or call you back right away!
Trestles LH femtosecond lasers with integrated DPSS DMPLH laser pump - DPSS DMP LH series advantages
Trestles LH10-fs/CW laser system at UC Santa Cruz Center of Nanoscale Optofluidics
Del Mar Photonics Tresltes LH laser used for STED microscopy of nanodiamons
Femtosecond Lasers
Trestles femtosecond Ti:Sapphire 
laser
Trestles Finesse femtosecond 
Ti:Sapphire laser with integrated DPSS 
pump laser
Trestles LH femtosecond Ti:Sapphire laser with integrated DPSS pump laser
Trestles Opus femtosecond Ti:Sapphire laser with built in 3 Watt DPSS laser
Teahupoo Rider femtosecond 
amplified Ti:Sapphire laser
Mavericks femtosecond 
Cr:Forsterite laser
Tamarack femtosecond fiber laser 
(Er-doped fiber)
Buccaneer femtosecond 
OA fiber laser (Er-doped fiber) and SHG
Cannon Ultra-broadband 
light source
Tourmaline femtosecond 
Yt-doped fiber laser
Yb-based high-energy fiber laser system kit, model Tourmaline Yb-ULRepRate-07
Ytterbium-doped Femtosecond Solid-State Laser Tourmaline Yb-SS400
From Professor Albert Diaspro
Dear friends,
I am proud to inform that our short review on multiphoton microscopy is the most 
read paper on biomedical engineering online:
http://www.biomedical-engineering-online.com/content/5/1/36.
All the best
Alby
ISTITUTO ITALIANO 
DI TECNOLOGIA
Prof. Alberto Diaspro
Scientific Head
Nanophysics
Via Morego, 30 16163 Genova
Tel: +39-010.71.781.503
Fax +39-010-72.03.21
Mobile +39-3666719968
www.iit.it
alberto.diaspro@iit.it
 
Multi-photon excitation microscopy
Alberto Diaspro1,2,5 , Paolo Bianchini1 , Giuseppe Vicidomini1 , 
Mario Faretta3 , Paola Ramoino4 and Cesare Usai5 
1 LAMBS-MicroScoBio Research Center, Department of Physics, University of Genoa, 
Via Dodecaneso 33, 16146 Genova, Italy
2 IFOM The FIRC Institute for Molecular Oncology Foundation, Via Adamello, 16, 
20139 Milan, Italy
3 IFOM-IEO Consortium for Oncogenomics European Institute of Oncology, via 
Ripamonti 435, 20141 Milan, Italy
4 DIPTERIS – Department for the Study of the Territory and its Resources, 
University of Genoa, Corso Europa 26, 16132 Genova, Italy
5 CNR- National Research Council, Institute of Biophysics, Via De Marini, 6, 
16149 Genova, Italy
author email corresponding author email
BioMedical Engineering OnLine 2006, 5:36doi:10.1186/1475-925X-5-36
The electronic version of this article is the complete one and can be found 
online at: http://www.biomedical-engineering-online.com/content/5/1/36
Received: 11 March 2006
Accepted: 6 June 2006
Published: 6 June 2006
© 2006 Diaspro et al; licensee BioMed Central Ltd.
This is an Open Access article distributed under the terms of the Creative 
Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which 
permits unrestricted use, distribution, and reproduction in any medium, provided 
the original work is properly cited.
Abstract
Multi-photon excitation (MPE) microscopy plays a growing role among 
microscopical techniques utilized for studying biological matter. In conjunction 
with confocal microscopy it can be considered the imaging workhorse of life 
science laboratories. Its roots can be found in a fundamental work written by 
Maria Goeppert Mayer more than 70 years ago. Nowadays, 2PE and MPE microscopes 
are expected to increase their impact in areas such biotechnology, neurobiology, 
embryology, tissue engineering, materials science where imaging can be coupled 
to the possibility of using the microscopes in an active way, too. As well, 2PE 
implementations in noninvasive optical bioscopy or laser-based treatments point 
out to the relevance in clinical applications. Here we report about some basic 
aspects related to the phenomenon, implications in three-dimensional imaging 
microscopy, practical aspects related to design and realization of MPE 
microscopes, and we only give a list of potential applications and variations on 
the theme in order to offer a starting point for advancing new applications and 
developments.
1. Introduction
There have been a variety of reasons for the continuing growth of interest in 
optical microscopy in spite of the low resolution with respect to modern 
scanning probe or electron microscopy [1]. The main reason lies in the fact that 
optical microscopy is still considered unique in allowing the 4D (x-y-z-t) 
examination of biological systems in a hydrated state in living samples or under 
experimental conditions that are very close to living or physiological states. 
This evidence coupled to fluorescence labelling and other advances in molecular 
biology permits to attack in an effective way the complex and delicate problem 
of the connection between structure and function in biological systems [2-6]. 
Within this framework, inventions in microscopy were stimulated, and contributed 
to the evolution of the optical microscope in its modern forms [7,8]. 
Multiphoton excitation microscopy is an important part of this progress in the 
field of microscopy applied to the study of biological matter from the 
inventions of the confocal microscope [9] and of the atomic force microscope 
[10]. Nowadays, confocal and multiphoton microscopes can be considered the 
imaging workhorses of life science laboratories [11].
Multiphoton excitation microscopy (MPE) has its roots in two-photon excitation 
(2PE) microscopy whose story dates back more than 70 years. In 1931, Maria 
Goeppert-Mayer published her brilliant doctoral dissertation on the theory of 
two-photon quantum transitions in atoms and established the theoretical basis 
behind 2PE [12]. This photophysical effect was experienced after the development 
of laser sources, as well, other non-linear related effects were also observed 
in the 60s and 70s [13,14]. In 1976, Berns reported about a probable two-photons 
effect as a result of focusing an intense pulsed laser beam onto chromosomes of 
living cells [15], and such interactions form the basis of modern nanosurgery 
[16] and targeted transfection [17]. One has to wait until 1978 to find the 
description of the first nonlinear scanning optical microscope with depth 
resolution. The reported microscopic imaging was based on second-harmonic 
generation (SHG) and the possibility of performing 2PE microscopy was outlined 
[18]. Unfortunately, applications in biology were hampered due by the high peak 
intensities required for priming 2PE fluorescence. This obstacle was overcome 
with the advent of ultrashort and fast pulsed lasers in the 80s [19]. Denk and 
colleagues in a seminal paper on 2PE laser scanning fluorescence microscopy 
clearly demonstrated the capability of 2PE microscopy for biology [20]. This 
fact brought a "new deal" in fluorescence microscopy [21-23]. Such a leap in 
scientific technology stimulated disparate disciplines and several variations on 
the theme extending studies from the tracking of individual molecules within 
living cells to the observation of whole organisms [24-28].
2. Foundations of 2PE microscopy
Two-photon excitation of fluorescent molecules is a non-linear process related 
to the simultaneous absorption of two photons whose total energy equals the 
energy required for the more familiar one-photon excitation (1PE) [29]. The 
excitation process of a fluorescent molecule under 1PE typically requires 
photons in the ultraviolet or blue/green spectral range. Under sufficiently 
intense illumination, usually provided by a laser source, the very same process, 
i.e. excitation of a fluorescent molecule from the ground to the excited state, 
can take place in the infrared spectral range. This situation is illustrated in 
figure 1 using a Perrin-Jablonski diagram: the sum of the energies of the "two" 
infrared photons colliding with the very same molecule has to be greater than 
the energy gap between the molecule's ground and excited states. Since 2PE 
excitation requires at least two statistically "independent" photons for each 
excitation process, its rate depends on the square power of the instantaneous 
intensity. 3PE and higher photon excitation is also possible. This implies that 
deep ultraviolet (UV) microscopy can be performed without having the 
disadvantages related to UV-matter interactions, i.e. polymerization or 
temperature effects. However, our treatment will be mainly conducted in terms of 
2PE for sake of simplicity but any step can be extended to MPE.
Figure 1. Perrin-Jablonski fluorescence diagram. Simplified Perrin-Jablonski 
scheme for 1PE, 2PE and 3PE. Once the excited state has been reached the 
subsequent fluorescent emission is the very same for the three different 
modalities of excitation.
The most popular relationship about 2PE is related to the practical situation of 
a train of beam pulses focused through a high numerical aperture (NA) objective, 
with a duration τp and fprepetition rate. The advantage of using a train of 
repeated beam pulses is given by the fact that high peak power can be used for 
priming the process resulting on an average power tolerated by the biological 
system under investigation. Under controlled conditions, the probability, na, 
that a certain fluorophore simultaneously absorbs two photons during a single 
pulse, in the paraxial approximation is given by [20]
where Pave is the average power of the illumination beam, δ2 is the two-photon 
cross section of the fluorescent molecule, and λ is the excitation wavelength. 
Introducing 1 GM (Goppert-Mayer) = 10-58 [m4 s], for a δ2 of approximately 10 
GM, focusing through an objective of NA >1, an average incident laser power of ≈ 
1–50 mW, operating at a wavelength ranging from 680 to 1100 nm with 80–150 fs 
pulsewidth and 80–100 MHz repetition rate, one should get fluorescence without 
saturation. It is convenient that for optimal fluorescence generation, the 
desirable repetition time of pulses should be on the order of typical 
excited-state lifetime, which is a few nanoseconds for commonly used fluorescent 
molecules. For this reason the typical repetition rate is around 100 MHz, i.e. 
one order of magnitude slower than typical fluorescence lifetime. As well, 
during the pulse time (10-13 s of duration and a typical lifetime in the 10-9 s 
range) the molecule has insufficient time to relax to the ground state. This can 
be considered a prerequisite for absorption of another photon pair. Therefore, 
whenever na approaches unity saturation effects start occurring. Such a 
consideration allows one to optimize optical and laser parameters in order to 
maximize the excitation efficiency without saturation. In case of saturation the 
resolution is declining and worsening the image. It is also evident that the 
optical parameter for enhancing the process in the focal plane is the lens 
numerical aperture, NA, even if the total fluorescence emitted is independent 
from this parameter. This value is usually kept around 1.3–1.4. As example, for 
fluorescein that possesses a δ2 ≅38 GM at 780 nm, using NA = 1.4, a repetition 
rate of 100 MHz and pulse width of 100 fs within a range of Pave assumed 1, 10, 
20 and 50 mW, from equation (1) one has na≈ 5930 (Pave)2. This means that, as 
function of the average excitation power 1, 10, 20, 50 mw one gets 5.93 10-3, 
5.93 10-1, 1.86, 2.965 respectively, with saturation starting at 10 mW. The 
related rate of photon emission per molecule, at a non saturation excitation 
level, in absence of photobleaching, is given by na multiplied by the repetition 
rate of the pulses, i.e. approximately 5·107 photons s-1. It is worth noting 
that when considering the effective fluorescence emission one should consider a 
further factor given by the so-called quantum efficiency of the fluorescent 
molecules. It is worth noting that the fluorophore emission spectrum results 
independent of the excitation mode from 1PE to MPE like the quantum efficiency.
Now, even if the quantum-mechanical selection rules for MPE differ from those 
for one-photon excitation, several common fluorescent molecules can be used. 
Unfortunately, the knowledge of 1PE cross-section for a specific fluorescent 
molecule does not allow any quantitative prediction of the two-photon trend. The 
only "rule of thumb" that one can use states that a 2PE cross-section peak can 
be expected at a 2 folds wavelength with respect to the 1PE case. Table 1 
summarises the excitation properties of some popular fluorescent molecules under 
2PE regime.
Table 1. 2PE excitation parameters.
3. Optical implications of 2PE
3.1 Optical sectioning and confocal imaging
The possibility of the three-dimensional reconstruction of the volume 
distribution of intensive parameters, as fluorescence emission, from biological 
systems is one of the most powerful properties of the optical microscope. To 
collect optical slices from a three-dimensional object the so-called optical 
sectioning [3,11] technique is used as depicted in figure 2. It is essentially 
based on a fine z stepping either of the objective or of the sample stage, 
coupled with the usual x-y image capturing. The synchronous x-y-z scanning 
allows the collection of a set of two-dimensional images, which are somehow 
affected by signal cross talk from other planes from the sample. In fact, the 
observed image Oj, obtainable when positioning the geometrical focus of the lens 
at a certain plane" j "within the specimen, is produced by the true fluorescence 
distribution Ij at plane "j", distorted by the microscope in some way that can 
be described by a function S, plus differently distorted contributions from 
adjacent "k" planes positioned above and below the actual plane, and noise N. 
Using a convenient and appropriate formalism one has:
Figure 2. Optical sectioning scheme. A three-dimensional sample can be sketched 
as a series of optical slices. Let's call slice j the one containing the 
geometrical focus of the objective and refer to the adjacent planes as k slices. 
The sample contains a three-dimensional distribution of fluorescently labelled 
molecules whose intensity distribution is I, slice by slice. The thickness of 
each optical slice is approximately one half of the axial resolution, say ≈λ/2.
Oj = IjSj + ∑k≠j IkSk + N. (2)
Equation (2) reflects the fact that when a set of two-dimensional images is 
acquired at various focus position and under certain conditions, in principle, 
one can recover the 3D shape of the object, described by the intensive parameter 
I, by solving the above set of equations and finding the best estimate for I, 
slice by slice. By this procedure, unwanted light can be computationally removed 
combining the image data from a stack of "k" images. Such an operation can be 
optically performed using some physical stratagems that are behind confocal and 
MPE/2PE scanning microscopy.
In confocal microscopy, the observed image is built up, scanning the sample 
point by point, by adding information from x-y-z sampled regions. The price to 
pay, or the main drawback, is that image formation is not as immediate as 
widefield techniques in which the whole image is acquired at the same time. This 
is due to the fact that in confocal microscopy the specimen is sequentially 
illuminated point by point and at the same time all is masked but the 
illuminated in-focus regions for providing return light to the detector. As 
shown in figure 3 an illumination and a detection pinhole are placed in the 
optical pathway. The detection pinhole – the mask – is placed in front of the 
detector at a plane that is conjugate to the in-focus or "j" plane, such that 
the illumination spot and the pinhole aperture are simultaneously focused at the 
same specimen volume. This coincidence of the illumination and detected volume 
is responsible for confocality. The final effect of such an optical 
implementation is that out-of-focus contributions are excluded from the detector 
surface and the observed image, Oj, is close to the true distribution of the 
fluorescence intensity I, plane by plane during the x-y-z scanning operations. 
This allows performing optical sectioning.
Figure 3. Confocal optical pathways. An illumination and a detection pinhole are 
placed in the optical pathway. The detection pinhole – the mask – is placed in 
front of the detector at a plane that is conjugate to the in-focus or "j" plane, 
such that the illumination spot and the pinhole aperture are simultaneously 
focused at the same specimen volume. This coincidence of the illumination and 
detected volume is responsible for confocality. The illumination pinhole allows 
to perform pointlike scanning.
3.2 The 2PE optical case
In terms of optical implications the two-photon effect has the important 
consequence of limiting the excitation region to within a sub-femtoliter volume. 
This means that the emission region is intrinsically confocal. The resulting 3D 
confinement in terms of image formation process can be described by means of 
consolidated optical considerations [30]. Using a certain excitation light at a 
wavelength λ, the intensity distribution within the focal region of an objective 
having numerical aperture NA = n sin (α) is given, in the paraxial regime, by
where J0 is the 0th order Bessel function, ρ is a radial coordinate in the pupil 
plane, n is the refractive index of the medium between the lens and the 
specimen, (α) is the semi-angle of aperture of the lens [31],
and
are dimensionless axial and radial coordinates, respectively, normalized to the 
wavelength. Now, the intensity of fluorescence distribution within the focal 
region has a I(u, v) behaviour for the 1PE case [31]. In case of 2PE one has to 
consider a double wavelength and a square behaviour, i.e. I2(u/2, v/2). As 
compared with the 1PE case, the 2PE emission intensity distribution is axially 
confined.
In fact, considering the integral over ν, keeping u constant, its behaviour is 
constant along z for one-photon and has a half-bell shape for 2PE. This 
behaviour is responsible of the 3D discrimination property of 2PE, i.e. of the 
optical sectioning properties of the 2PE microscope.
Now, the most interesting aspect is that the excitation power falls off as the 
square of the distance from the lens focal point, within the approximation of a 
conical illumination geometry [31]. In practice this means that the square 
relationship between the excitation power and the fluorescence intensity brings 
about the fact that 2PE falls off as the fourth power of distance from the focal 
point of the objective. This fact implies that those molecules away from the 
focal region of the objective lens do not contribute to the image formation 
process and are not affected by photobleaching or phototoxicity. Since these 
molecules are not involved in the excitation process, a confocal-like effect is 
obtained without the necessity of a confocal pinhole. It is immediately evident 
that in this case the optical sectioning effect is obtained in a physically 
different way with respect to the confocal case. Accordingly the optical set-up 
is simplified, under some aspects, see figure 4.
Figure 4. MPE simplified optical pathways. In the MPE optical pathways the 
emission pinhole is removed since the only emitted light reaching the sensor is 
coming from the currently point scanned volume in the sample. No other 
fluorescence signal is produced elsewhere.
Figure 5 and figure 6 show the differences in terms of excitation-emission 
process between confocal and multiphoton schemes, respectively. The consequences 
of the spatial confinement of the MPE result in a consequent confinement of the 
emitted fluorescence: in the confocal case all the molecules within the double 
cone of excitation are involved in the light-matter interaction while in the MPE 
case such interaction is restricted to a small volume centred at the geometrical 
focus of the objective. The immediate consequence is that a 2PE microscope is an 
intrinsically three-dimensional image formation system. This fact has also very 
important consequences on the photobleaching processes. So far, in the 2PE case 
no fluorescence has to be removed from the detection pathway since fluorescence 
can exclusively come from a small focal volume that has a capacity of the order 
of fraction of femtoliter. In fact, in 2PE over 80% of the total intensity of 
fluorescence comes from a 700–1000 nm thick region about the focal point for 
objectives with numerical apertures in the range from 1.2 to 1.4 [30]. The fact 
that the background signal coming from adjacent planes tends to zero produces a 
sort of compensation for the reduced spatial resolution due to the utilization 
of a wavelength that is twice with respect to the 1PE case, as shown in figure 
7. On the other hand, the utilisation of infrared wavelengths instead of 
UV-visible ones allows achieving a deeper penetration than in conventional case 
[32]. This is due to the fact that the scattering effect is proportional to the 
inverse fourth power of the wavelength. Thus the longer wavelengths used in 2PE, 
or in general in MPE, will be scattered less than the ultraviolet-visible 
wavelengths used for conventional excitation allowing to reach fluorescence 
targets in depth within thick samples (approx. 1 mm). It has been shown that 
two-photon fluorescence images can be obtained throughout almost the entire grey 
matter of the mouse neocortex by using optically amplified femtosecond pulses. 
The achieved imaging depth approaches the theoretical limit set by excitation of 
out-of-focus fluorescence [33]. The fluorescence light emitted, on the way back, 
can be more efficiently collected using a large area detector since it can 
uniquely come from the sub-femtoliter 2PE volume of event.
Figure 5. Confocal fluorescence emission distribution. The emission process, in 
green, under blue 1PE excitation is broadened in the whole double cone 
excitation shape within the analyzed cell.
Figure 6. MPE fluorescence emission distribution. Confinement of the emission 
process, in green, under red 2PE. Under 2PE the only molecules excited are those 
confined in a small subfemtoliter volume at the illumination beam focus 
position. This is particularly relevant for the photobleaching process that is 
globally reduced with respect to the 1PE case. The capacity of the volume can be 
approximatively calculated by using the resolution parameters of the optical 
system, since they are indicators of the volume containing the maximum photon 
density. This is valid only for non saturated processes. Under saturation of 
fluorescence beam intensity plays an important role in the emission shape and 
optical resolution.
Figure 7. Pointlike emitter optical response. From left to right: calculated x-y 
(above) and r-z (below) intensity distributions, in logarithmic scale, for a 
point like source imaged by means of wide-field, 2PE and confocal microscopy. 
Both 2PE and confocal shapes exhibit a better signal to noise ratio than 
widefield case. 2PE distribution is larger due the fact that a wavelength twice 
than in the wide-field and confocal case is responsible for the intensity 
distribution. Such intensity distributions are also known as point spread 
functions of the related microscopes. Optical conditions: excitation wavelengths 
are 488 nm and 900 nm for 1PE and 2PE, respectively; emission wavelength is 520 
nm; numerical aperture is 1.3 for an oil immersion objective with oil refractive 
index value set at 1.515.
4. Practical aspects for the realization of a 2PE microscope
The main elements to realize a 2PE/MPE architecture, including confocal 
modality, are the following: high peak-power laser delivering moderate average 
power (fs or ps pulsed at relatively high repetition rate) emitting infrared or 
near infrared wavelengths (650–1100 nm), laser sources for confocal 1PE, a laser 
beam scanning system, high numerical aperture objectives (>1), a high-throughput 
microscope pathway, a spectral separation module for the emitted signal 
discrimination, and a high-sensitivity detection system [34]. Figure 8 shows a 
general scheme for a MPE microscope also illustrating two popular approaches 
that can be used for image formation, namely: de-scanned and non de-scanned 
mode. The former uses the very same optical pathway and mechanism employed in 
confocal laser scanning microscopy. The latter mainly optimises the optical 
pathway by minimising the number of optical elements encountered on the way from 
the sample to detectors, and increases the detector area. MPE non-descanned mode 
allows very good performances providing superior signal-to-noise ratio inside 
strongly scattering samples [32,33]. In the de-scanned approach pinholes are 
removed or set to their maximum aperture and the emission signal is captured 
using the very same optical scanning pathway used for excitation. In the latter, 
the aim is to optimize the collection efficiency: pinholes are removed and the 
radiation emitted without passing through the laser beam scanning mirrors. 
Photomultiplier tubes are the most popular detectors in MPE microscopy. 
Avalanche photodiodes are also excellent in terms of sensitivity exhibiting 
quantum efficiency close to 70%–80% in the visible spectral range. Unfortunately 
they are high cost and the small active photosensitive area could introduce 
drawbacks in the detection scheme and require special de-scanning optics. CCD 
cameras are generally used in video rate multifocal imaging. Laser sources 
represent the core element for the 2PE/MPE microscope since for MPE high photon 
flux densities are required, > 1024 photons cm-2s-1. Using radiation in the 
spectral range of 650–1100 nm for MPE, excitation intensities in the MW-GW cm-2 
have to be produced. Nowadays, laser sources suitable for 2PE can be described 
as "turnkey" systems, and Ti Sapphire lasers are the most utilized due to the 
high coincidence with the 2PE wavelengths needed for the majority of the 
commonly used fluorescent molecules. Other laser sources used for 2PE are 
Cr-LiSAF, pulse-compressed Nd-YLF in the femtosecond regime, and mode-locked 
Nd-YAG and picosecond Ti-Sapphire lasers in picosecond regime. Moreover the 
absorption coefficients of most biological samples, cells and tissues are 
minimised within this spectral window. Table 2 reports data on the currently 
available laser sources for applications in MPE microscopy and spectroscopy. The 
parameters that are more relevant in the selection of the laser source are 
average power, pulsewidth and repetition rate, and wavelength also accordingly 
to equation (1). The most popular features for an infrared pulsed laser are 
700mW-1W average power, 80–100 MHz repetition rate, and 100–150 fs pulse width. 
So far, the use of short pulses and high repetition rates are mandatory to allow 
image acquisition in a reasonable time while using power levels that are 
biologically tolerable. In order to minimise pulse width dispersion problems it 
should be considered to operate with pulses around 150 nm. This constitutes a 
very good compromise both for pulse stretching and sample viability. One should 
always remind that a shorter pulse broadens more than a longer one. Advances in 
laser sources are going on considering more compact sources, large tunability 
range, high average power, and special designs for tailored needs at lower 
prices [35]. Objective lenses influence the performances of any optical 
microscope, and for a MPE system special considerations have to be done taking 
into the proper account both equations (1) and (3). New technological requisites 
have to be considered with respect to conventional excitation fluorescence 
microscope. An adequate transmission in the IR regime has to be coupled with 
good collection efficiency towards the ultraviolet region. Moreover, the number 
of components should be minimised without affecting resolution properties in 
order to reduce pulse widths distortions. Although the collection efficiency of 
the time averaged photon flux is dependent on the numerical aperture of the 
collecting lens, the total fluorescence generation is independent of the 
numerical aperture of the focusing lens when imaging thick samples. Figure 9 
shows an example of optical sectioning performed through a thick sample that 
exploits the autofluorescence signal. This is due to the fact that the increase 
of intensity, obtained by a sharper focusing (high NA), is counterbalanced by 
the shrinking of the excitation volume. Thus the total amount of fluorescence 
summed over the entire space remains constant. The very relevant practical 
consequence of this fact is that in 2PE measurements on thick samples, assuming 
no aberrations, the generated fluorescence is insensitive to the size of the 
focal spot. As a positive consequence, a moderate variation of the laser beam 
size would not affect the measurements. This is a very efficient condition due 
to the fact that using an appropriate (non-descanned) acquisition scheme it is 
possible to collect all the generated fluorescence. In terms of pulse 
broadening, a 100 fs pulse can result between 1.14 and 1.23 times at the sample 
using a good lens. Once a MPE architecture has been realized, one should 
consider to keep under control the following parameters: power and pulse width 
at the sample focal plane checking for the square intensity/power behaviour, 
spectral separation of the emitted fluorescence including removal of the 
possible excitation reflections that could be particularly subtle, z-axis 
precise control and laser-scanning system alignment [34]. Figure 10 and figure 
11 demonstrate the multiple fluorescence imaging capability and 3D multiple 
fluorescence imaging, respectively.
Figure 8. MPE simplified optical schemes. Descanned (red dot) and non-descanned 
(blue dot) schemes for 2PE microscopy. Due to the confinement of the excitation 
process when operating in non-descanned mode the collection efficiency in depth 
is increased as shown by the two side views from the very same thick scattering 
sample (Courtesy of Mark Cannel and Christian Soeller; image inset courtesy of 
Ammasi Periasamy).
Table 2. MPE laser table.
Figure 9. Optical sectioning using 2PE autofluorescence. 2PE optical sectioning 
of Colpoda maupasi resting cysts, 21–32 μm average dimensions. Encystment is 
particularly widespread in species living in ephemeral fresh-water puddles and 
is induced by exhaustion of the food and drying out. The resulting images have 
been obtained at LAMBS-MicroScoBio exploiting autofluorescence primed using 740 
nm excitation. A Chameleon-XR ultrafast Ti-Sapphire laser (Coherent Inc., USA) 
and a Nikon PCM2000 confocal scanning head have been used [34]. Linear frame 
dimension is 70 μm, z steps have been performed every 0.5 μm, and not all the 
optical sections are imaged.
Figure 10. Multiple fluorescence 2PE imaging. 2PE multiple fluorescence image 
from a 16 μm cryostat section of mouse intestine stained with a combination of 
fluorescent stains (F-24631, Molecular Probes). Alexa Fluor 350 wheat germ 
agglutinin, a blue-fluorescent lectin, was used to stain the mucus of goblet 
cells. The filamentous actin prevalent in the brush border was stained with 
red-fluorescent Alexa Flu or 568 phalloidin. Finally, the nuclei were stained 
with SYTOX ® Green nucleic acid stain. Imaging has been performed at 780 nm, 100 
x 1.4 NA Leica objective, using a Chameleon XR ultrafast Ti-Sapphire laser 
(Coherent Inc., USA) coupled at LAMBS-MicroScoBio with a Spectral Confocal Laser 
Scanning Microscope, Leica SP2-AOBS.
Figure 11. 3D and 2D fluorescence projections. Pictorial representation of the 
3D and 2D projections of multiple fluorescence from a marine sponge, Chondrilla 
nucula. The specimen has been loaded with Alexa 488 fluorescent molecules 
specific aminobutirric acid (GABA) emitting in green, DAPI for nuclear DNA for 
the blu component. Red signals are due to the autofluorescence from symbiontic 
bacteria contamination. Imaging has been perfomed using a Chameleon XR ultrafast 
Ti-Sapphire laser (Coherent Inc., USA) coupled at LAMBS-MicroScoBio with a 
Spectral Confocal Laser Scanning Microscope, Leica SP2-AOBS. (Sample 
availability and preparation, courtesy of Renata Manconi, University of Sassari, 
Roberto Pronzato and Lorenzo Gallus, University of Genoa).
5. Application trends and conclusions
2PE and MPE microscopes are expected to increase their impact in areas such 
biotechnology, neurobiology, embryology, tissue engineering, materials science 
where imaging can be coupled to the possibility of using the microscopes in an 
active way, too. Clinically, 2PE may find applications in non-invasive optical 
bioscopy, while in cell biology the imaging abilities are coupled to the 
possibility of producing localized chemical reactions. Potential applications to 
integrative cardiac physiology or the possibility of tracking for long time 
biological events in living systems point out to the ability of making direct 
observations of phenomena and circumstances that before could only be inferred 
using other approaches. The myriad of new investigation possibilities offered by 
2PE/MPE microscopy enlarges so much the fields of application that it is not 
possible to outline in a complete way all the variations that can take place. 
For this reason, we give in this last paragraph asummary of main properties of 
MPE and a limited overview of paramount trends.
The great impact of 2PE in optical microscopy is related to the fact that it 
couples a three-dimensional intrinsic ability with almost five other interesting 
capabilities [21]. First, 2PE greatly reduces photo-interactions and allows 
imaging of living specimens on long time periods. Second, it allows operating in 
a high-sensitivity background-free acquisition scheme. Third, 2PE microscopy can 
penetrate turbid and thick specimens down to a depth of a few hundreds 
micrometers. Fourth, due to the distinct character of the multiphoton absorption 
spectra of many of the fluorophores 2PE allows simultaneous excitation of 
different fluorescent molecules reducing colocalization errors. Fifth, 2PE can 
prime photochemical reactions within subfemtoliter volumes inside solutions, 
cells and tissues.
Moreover, the advances in the field of fluorescent markers added value and 
potential to MPE microscopy. It is worth mentioning: the design of application 
suited chromophores [36]; the development and utilization of the so-called 
quantum dots [37]; the use of visible [38] and photoactivatable [39,40] 
fluorescent proteins from the green fluorescent protein (GFP) and its natural 
homologues to specifically engineered variants [6], the use of photoswitchable 
proteins to break the diffraction barrier in fluorescence microscopy at low 
light intensities [41].
Furthermore, this form of non-linear microscopy also supported the development 
and application of several investigation techniques, among them: three-photon 
excited fluorescence [42], second harmonic generation [43], third-harmonic 
generation [44], fluorescence correlation spectroscopy [45], image correlation 
spectroscopy [46], single molecule detection [47,48]; photodynamic therapies 
[49], and flow cytometry [50]. There is also an ongoing research activity to use 
2PE and MPE in new fields where its special features can be advantageously 
applied to improve and to optimise existing schemes [51]. This covers new online 
detection systems like endoscopic imaging based on gradient refractive index 
fibres [52], the development of new substrates with higher fluorescence output 
[53] as well as the use of 2PE to systematically crosslink protein matrices and 
control the diffusion [54] and to perform localized uncaging [55]. When 
considering the growing interest for detection/sensing technology in medical 
diagnostics and biotechnology, one should not ignore the recent explosion in the 
use of metallic nanostructures to favourably modify the spectral properties of 
fluorophores and to alleviate some fluorophore photophysical constraints. Within 
the framework the fusion of MPE with metal-enhanced fluorescence has a powerful 
potential in biotechnology: from immunoassay to enhanced ratiometric sensing and 
DNA detection [56]. A further mention is due to biomolecular tracking in real 
time and in vivo. Here 2PE and MPE can be considered as the dominant 
technologies. One mention is for in vivo brain imaging realized by means of ea 
newly designed compact and portable 2PE micro endoscope recently used to 
visualize hippocampal blood vessels in the brains of live mice [57]. As well a 
first partial view into the dynamics of developmentally programmed, long-range 
cell migration in the mammalian thymus was obtained using in a 4D (x-y-z-t) 
manner 2PE. So, the movement of thymocytes was followed in real time through the 
cortex within intact thymic lobes [58]. All these facts point out to the 
consideration that MPE makes possible to perform a 7D exploration of living 
cells due to its inherent ability in (x-y-z-t), FLIM (Fluorescence Lifetime 
Imaging Microscopy), FRAP (Fluorescence Recovery After Photobleaching), FRET 
(Forster-fluorescence Resonance Energy Transfer) and SHG (Second Harmonic 
Generation) [24]. Additionally, regardless of the fact that all far field light 
microscopes are limited in the achievable diffraction-limited resolution, MPE is 
pushing modern light microscopy towards fluorescence optical nanoscopy [59,60].
Table 3. 2PE vs. 1PE optical microscopy.
Acknowledgements
The authors dedicate this work to Osamu Nakamura, a pioneer in 2PE, who passed 
away on 23 Jan 2005 at Handai Hospital, Japan. AD dedicates this work to Mario 
Arace and is still using his oscilloscope, purchased in 1978, in the lab. The 
first Italian 2PE architecture realized at LAMBS of the University of Genoa has 
been supported by INFM grants. LAMBS-MicroScoBio is currently granted by IFOM 
(Istituto FIRC di Oncologia Molecolare, FIRC Institute of Molecular Oncology, 
Milan, Italy) and Fondazione San Paolo (Torino, Italy). LAMBS joins the NANOMED 
Italian Research Program. MicroScoBio is a Research Center of the University of 
Genoa on Correlative Microscopy and Spectroscopy with applications in 
Biomedicine and Oncology, Genoa, Italy. IFOM is the FIRC Foundation on Molecular 
Oncology, Milan, Italy. LAMBS is a multicenter Laboratory for Advanced 
Microscopy, Bioimaging and Spectroscopy http://www.lambs.it.
References
Diaspro A: New World Microscopy.
IEEE Engineering In Medicine And Biology Magazine 1996 , 15:29-100. Publisher 
Full Text 
Beltrame F, Bianco B, Castellaro G, Diaspro A: Fluorescence, Absorption, 
Phase-contrast, Holographic and Acoustical Cytometries of Living Cells. In 
Interactions between Electromagnetic Fields and Cells. Volume 97. Edited by: 
Chiabrera A. NATO ASI Series, Plenum Press Publishing, New York and London; 
1985:483-498. 
Arndt-Jovin DJ, Nicoud RM, Kaufmann J, Jovin TM: Fluorescence digital-imaging 
microscopy in cell biology.
Science 1985 , 230:13330-13335. 
Fay FS, Carrington W, Fogarty KE: Three-dimensional molecular distribution in 
single cells analyzed using the digital imaging microscope.
J Microsc 1989 , 153:133-149. PubMed Abstract 
Periasamy A: Methods in Cellular Imaging. Oxford University Press, New York; 
2000.
Zhang J, Campbell RE, Ting A, Tsien RY: Creating new fluorescent probes for cell 
biology.
Nature Review Molecular Biology 2002 , 3:906-918. Publisher Full Text 
Amos B: Lessons from the history of light microscopy.
Nat Cell Biol 2000 , 2(8):E151-152. PubMed Abstract | Publisher Full Text 
Bastiaens PI, Hell SW: Recent Advances in Light Microscopy.
Journal of Structural Biology 2004 , 147:1-89. Publisher Full Text 
Pawley JB: Handbook of Biological Confocal Microscopy. 3rd edition. Plenum 
Press- Springer, New York; 2006.PubMed Abstract | Publisher Full Text 
Jena BP, Horber JHK: Atomic Force Microscopy in Cell Biology. In Methods in Cell 
Biology. Volume 68. Academic Press; 2002.PubMed Abstract 
Amos WB, White JG: How the Confocal Laser Scanning Microscope entered Biological 
Research.
Biology of the Cell 2003 , 95:335-342. PubMed Abstract | Publisher Full Text 
Göppert-Mayer M: Über Elementarakte mit zwei Quantensprüngen.
Ann Phys 1931 , 9:273-295. 
Masters BR, So PT: Antecedents of two-photon excitation laser scanning 
microscopy.
Microsc Res Tech 2004 , 63:3-11. PubMed Abstract | Publisher Full Text 
Esposito A, Federici F, Usai C, Cannone F, Chirico G, Collini M, Diaspro A: 
Notes on theory and experimental conditions behind two-photon excitation 
microscopy.
Microsc Res Tech 2004 , 63:12-17. PubMed Abstract | Publisher Full Text 
Berns MW: A possible two-photon effect in vitro using a focused laser beam.
Biophys J 1976 , 16:973-977. PubMed Abstract | PubMed Central Full Text 
Konig K: Multiphoton microscopy in life sciences.
J Microsc 2000 , 200:83-104. PubMed Abstract | Publisher Full Text 
Tirlapur UK, Konig K: Cell biology: Targeted transfection by femtosecond laser.
Nature 2002 , 418:290-291. PubMed Abstract | Publisher Full Text 
Sheppard CJ, Kompfner R: Resonant scanning optical microscope.
Appl Opt 1978 , 17:2879-2885. 
Girkin JM: Optical physics enables advances in multiphoton imaging.
Journal of Physics D: Applied Physics 2003 , 36:R250-R258. Publisher Full Text
Denk W, Strickler JH, Webb WWW: Two-photon laser scanning fluorescence 
microscopy.
Science 1990 , 248:73-76. PubMed Abstract 
Diaspro A: Confocal and two-photon microscopy : foundations, applications, and 
advances. Wiley-Liss, New York; 2002.
Pennisi E: Biochemistry: Photons Add Up to Better Microscopy.
Science 1999 , 275:480-481. Publisher Full Text 
Diaspro A: Rapid dissemination of two-photon excitation microscopy prompts new 
applications.
Microsc Res Tech 2004 , 63:1-2. PubMed Abstract | Publisher Full Text 
Zoumi A, Yeh A, Tromberg BJ: Imaging cells and extracellular matrix in vivo by 
using second-harmonic generation and two-photon excited fluorescence.
Proc Natl Acad Sciences (USA) 2002 , 99:11014-11019. Publisher Full Text 
Zipfel WR, Williams RM, Webb WW: Nonlinear magic: multiphoton microscopy in the 
biosciences.
Nature Biotechnology 2003 , 21:1369-1377. PubMed Abstract | Publisher Full Text
Helmchen F, Denk W: Deep tissue two-photon microscopy.
Nat Methods 2005 , 2:932-940. PubMed Abstract | Publisher Full Text 
Rubart M: Two-Photon Microscopy of Cells and Tissue.
Circ Res 2004 , 95:1154-1166. PubMed Abstract | Publisher Full Text 
Cruz HG, Lüscher C: Applications of two-photon microscopy in the neurosciences.
Frontiers in Bioscience 2005 , 10:2263-2278. PubMed Abstract | Publisher Full 
Text 
Callis PR: Two-photon-induced fluorescence.
Ann Rev Phys Chem 1997 , 48:271-297. Publisher Full Text 
Nakamura O: Three-dimensional imaging characteristics of laser scan fluorescence 
microscopy: Two-photon excitation vs. single-photon excitation.
Optik 1993 , 93:39-42. 
Bianco B, Diaspro A: Analysis of the three dimensional cell imaging obtained 
with optical microscopy techniques based on defocusing.
Cell Biophys 1989 , 15:189-200. PubMed Abstract 
Centonze VE, White JG: Multiphoton excitation provides optical sections from 
deeper within scattering specimens than confocal imaging.
Biophys J 1998 , 75:2015-2024. PubMed Abstract | Publisher Full Text | PubMed 
Central Full Text 
Theer P, Hasan MT, Denk W: Two-photon imaging to a depth of 1000 μm in living 
brains by use of a Ti:Al2O3 regenerative amplifier.
Optics Lett 2003 , 28:1022-1024. 
Diaspro A: Building a two-photon microscope using a laser scanning confocal 
architecture. In Methods in Cellular Imaging. Edited by: Periasamy A. Oxford 
University Press, New York; 2001:162-179. 
Girkin J, McConnell G: Advances in Laser Sources for Confocal and Multiphoton 
Microscopy.
Microsc Res Tech 2005 , 67:8-14. PubMed Abstract | Publisher Full Text 
Abbotto A, et al.: Dimethyl-pepep: a DNA probe in two-photon excitation cellular 
imaging.
Biophys Chem 2005 , 114: :35-41. PubMed Abstract | Publisher Full Text 
Jaiswal JK, Simon S: Potentials and pitfalls of fluorescent quantum dots for 
biological imaging.
Trends Cell Biol 2004 , 14:497-504. PubMed Abstract | Publisher Full Text 
Matz MV, Lukyanov KA, Lukyanov SA: Family of the green fluorescent protein: 
journey to the end of the rainbow.
Bioessays 2002 , 24:953-959. PubMed Abstract | Publisher Full Text 
Schneider M, Barozzi S, Testa I, Faretta M, Diaspro A: Two-photon activation and 
excitation properties of PA-GFP in the 720–920-nm region.
Biophys J 2005 , 89:1346-1352. PubMed Abstract | Publisher Full Text 
Post JN, Lidke KA, Rieger B, Arndt-Jovin DJ: One- and two-photon photoactivation 
of a paGFP-fusion protein in live Drosophila embryos.
FEBS Lett 2005 , 579:325-330. PubMed Abstract | Publisher Full Text 
Hofmann M, Eggeling C, Jakobs S, Hell SW: Breaking the diffraction barrier in 
fluorescence microscopy at low light intensities by using reversibly 
photoswitchable proteins.
Proc Natl Acad Sci U S A 2005 , 102:17565-17569. PubMed Abstract | Publisher 
Full Text | PubMed Central Full Text 
Hell SW, Bahlmann K, Schrader M, Soini A, Malak H, Gryczynski I, Lakowicz JR: 
Three-photon excitation in fluorescence microscopy.
J Biomedical Optics 1996 , 1:71-74. Publisher Full Text 
Campagnola PJ, Loew LM: Second-harmonic imaging microscopy for visualizing 
biomolecular arrays in cells, tissues and organisms.
Nat Biotechnol 2003 , 21:1356-1360. PubMed Abstract | Publisher Full Text 
Mueller M, Squier J, Wilson KR, Brakenhoff GJ: 3D microscopy of trasparent 
objects using third-harmonic generation.
J Microsc 1998 , 191:266-274. PubMed Abstract | Publisher Full Text 
Schwille P: Fluorescence correlation spectroscopy and its potential for 
intracellular applications.
Cell Biochem Biophys 2001 , 34:383-408. PubMed Abstract | Publisher Full Text
Wiseman PW, et al.: Spatial mapping of integrin interactions and dynamics during 
cell migration by image correlation microscopy.
J Cell Sci 2004 , 117:5521-5534. PubMed Abstract | Publisher Full Text 
Sonnleitner M, Schutz GJ, Schmidt T: Imaging individual molecules by two-photon 
excitation.
Chem Phys Lett 1999 , 300:221-226. Publisher Full Text 
Chirico G, Cannone F, Beretta S, Diaspro A: Single molecule studies by means of 
the two-photon fluorescence distribution.
Microsc Res Techniq 2001 , 55:359-364. Publisher Full Text 
Bhawalkar JD, Kumar ND, Zhao CF, Prasad PN: Two-photon photodynamic therapy.
J Clin Laser Med Surg 1997 , 15:201-204. PubMed Abstract 
Diaspro A: Two-photon fluorescence excitation. A new potential perspective in 
flow cytometry.
Minerva Biotecnologica 1998 , 11:87-92. 
McConnell G, Riis E: Two-photon laser scanning fluorescence microscopy using 
photonic crystal fibre.
J Biomed Opt 2004 , 9:922-927. PubMed Abstract | Publisher Full Text 
Jung JC, et al.: In vivo mammalian brain imaging using one- and two-photon 
fluorescence microendoscopy.
J Neurophysiol 2004 , 92:3121-3133. PubMed Abstract | Publisher Full Text 
Kappel C, et al.: Giant enhancement of two-photon fluorescence induced by 
resonant double grating waveguide structures.
Applied Physics B-Lasers and Optics 2004 , 79:531-534. Publisher Full Text 
Basu S, Campagnola PJ: Properties of crosslinked protein matrices for tissue 
engineering applications synthesized by multiphoton excitation.
J Biomed Mater Res A 2004 , 71:359-368. PubMed Abstract | Publisher Full Text
Diaspro A, et al.: Two-Photon Photolysis of 2-Nitrobenzaldehyde Monitored by 
Fluorescent-Labeled Nanocapsules.
J Phys Chem B 2003 , 107:11008-11012. Publisher Full Text 
Aslan K, et al.: Metal-enhanced fluorescence: an emerging tool in biotechnology.
Curr Opin Biotechnol 2005 , 16:55-62. PubMed Abstract | Publisher Full Text 
Flusberg BA, et al.: In vivo brain imaging using a portable 3.9 gram two-photon 
fluorescence microendoscope.
Opt Lett 2005 , 30:2272-2274. PubMed Abstract | Publisher Full Text 
Witt CM, et al.: Directed migration of positively selected thymocytes visualized 
in real time.
PLoS Biol 2005 , 3:e160. PubMed Abstract | Publisher Full Text | PubMed Central 
Full Text 
Hell SW: Toward fluorescence nanoscopy.
Nat Biotechnol 2003 , 21:1347-1355. PubMed Abstract | Publisher Full Text 
Egner A, Hell SW: Fluorescence microscopy with super-resolved optical sections.
Trends Cell Biol 2005 , 15:207-215. PubMed Abstract | Publisher Full Text